by Record-changer »
Thu Jan 06, 2011 6:34 am
It's the records.
It's called a groove guard, and is the reason Magnavox replaced the changer that can play odd sized records with changers that play only the 3 standard sizes.
They changed the design of the records, to the chagrin of the changer manufacturers. The record is thinner between the groove guard outer ridge and the label, so the grooves do not touch each other. Unfortunately, this causes havoc with older changers, because the stylus sets on the groove guard, rather than inside it. Gravity then causes the arm to slide inward, jumping the stylus into the music.
Newer changers have the 12" point set slightly farther in, but this can't be done on the changers that do odd sized records without also affecting the 7" set down point.
If you play few 7" records, you can change the setdown point adjustment with the screw in the hole in the plinth between the arm shaft and the turntable. Adjust it to land inside the groove guard.
I have been trying to design something to prevent this problem, but so far have not been completely successful.
One thing I did that helped some was to slow down the change cycle. This causes the arm to lower slower. I readjusted the change cycle idler height to engage the 33 step instead of the 45 step. I then adjusted the position of the motor assembly slightly so it would drive the cycle.
If this is happening on records without groove guards, then other effects could be occurring:
1. Skating, because the arm has overhang. Since this arm has no antiskating, the pickup drag could cause the stylus to move toward the spindle. It is worse with higher tracking forces.
2. A bent or loose locator leaf spring on the change cycle cam. This pinches the pickup arm positioning finger between the outer cam wall and the spring, so the setdown is consistent. If it is bent, it can flip the positioning finger to one side as the arm sets down.
Note: The locator spring affects 7" records too.
3. The arm height setscrews are misadjusted. There is a screw on top of the arm just above the arm shaft, and another screw at the right rear of the plinth. These need adjustment if the arm suddenly drops to the record at the end of the change cycle after lowering slower.
NOTE: If the record plays properly if it is higher in the stack, then the height adjustment is the trouble.
4. The pickup arm positioning finger is set too high off the cam face. This causes it to jump over the locator spring. The adjustment screw and nut are on the opposite side of the friction clutch from the positioning finger. Adjust for 1/32 inch between the finger and the cam face when the changer is out of cycle.
Note: The pickup positioning lever adjustment affects 7" records too.
5. Foreign matter on the cam face. There should be very little lubrication on it. Too much can drag the positioning finger to one side. It can also prevent the stylus from tracking the groove. I use graphite cylinder lock lube on mine.